Downloads


    Downloads are not available for this product.

The ROTAFLOW Centrifugal Pump System offers state-of-the-art centrifugal pump technology for best performance and maximum safety.

The ROTAFLOW Centrifugal Pump System is a natural part of the HL 20 Heart-Lung Machine and can also be used as a fully independent stand-alone unit.

ROTAFLOW Console

  • The ROTAFLOW Console operates independently as stand-alone with its own battery back-up and power supply.
  • It also perfectly complements the HL 20 product line.
  • Its compact design includes display for all vital information at a glance.
  • The unique flow and bubble measuring function is an integral part of the ROTAFLOW system.
  • Its compact design includes display for all vital information at a glance.

Integrated into the Heart-Lung Machine HL 20

  • The ROTAFLOW Console as HL 20 pump module, fully integrated into the Heart-Lung Machine HL 20.
  • In this configuration, all HL 20 supervising units will control and monitor parameters like pressure, bubble and low-level sensors.
  • The centrifugal pump operates quietly and smoothly on safe, low-voltage power supplied by the console.
  • There are no additional cables to connect.

Compact in size and low in weight

  • The ROTAFLOW Console is flexible: it can easily be removed from and reinstalled onto the HL 20, simply by lifting and positioning the unit according to guide pins.

Additional Features as HL 20 Pump Module:

  • Controlled by all supervising units (bubble, pressure, level)
  • Pulsatile flow mode option

One cable - three functions

  • The unique set-up of the ROTAFLOW Pump Drive also incorporates flow sensor and bubble detection into one advanced unit.
  • The blood flow is precisely measured in the pump outlet by ultrasonic transit time technology.
  • Costly disposable flow probes and time-consuming tubing connections are a thing of the past.

Optimum flexibility

  • The swivel-joint bracket assures ideal positioning with greater flexibility in handling, and allows shorter tubing lengths.
  • For further convenience, this sturdy drive's special cooling system has no dirtcatchingribs: a smooth surface for quick and easy cleaning.

The ROTAFLOW Emergency Drive

  • The emergency drive is designed for secure handling and fast fixation on the mast when every second counts.
  • While operating in manual mode, the LED displays the rpm
Safety
Certificate CE 0413
Standards IEC 60601-1, Type B, Class I
EMI EN 60601-1-2
Classification (MDD, Annex IX) IIb
Water protection IP64
Maximum flow deviation in case of error +20 %
Speed
Range 0 to 5000 RPM
Display resolution 1 RPM
Display accuracy ±20 RPM
Flow
Range 0 to 9.99 LPM
Display resolution 0.01 LPM
Display accuracy 0 to 1 LPM 0.07 LPM + offset drift
Display accuracy 1 to 9.99 LPM 7 % of actual value + offset drift
Offset drift max. 0.05 LPM
Environmental conditions
  Operation Storage Transportation
Temperature 17°C to 40°C -18°C to 45°C -18°C to 55°C
Relative humidity (non condensing) 30 % to 75 % 10 % to 96 % 10 % to 96 %
Air pressure 87 to 106 kPa 70 to 106 kPa 70 to 106 kPa

ROTAFLOW Console

Independent power supply, stand-alone
Mains (factory selectable) 100 V~ / 115 V~ / 230 V~ / 240 V~; 50 Hz / 60 Hz
Fuse 2 A for 100 V and 115 V 1 A for 230 V and 240 V
Power supply, Heart Lung Machine console
Voltage +24 V
Current max. 8 A
Batteries
Type NiCd
Voltage 24 V
Capacity 5 Ah
Charge/discharge cycles min. 500
Shelf life max. 6 months (under storage conditions)
Serial data output
Transfer rate 9600 baud
Protocol 8 N 1
Other
Measurements, L×W×H 179 × 385 × 243 mm
Weight ca. 14.4 kg

ROTAFLOW Drive Unit

Other
Coupling magnetic
Speed 0 to 5000 RPM
Pressure max. 750 mmHg
Flow meter ultrasonic
Bubble sensor ultrasonic
Pump stop air bubbles > 5 mm (0.065 cm3)
Motor brushless DC
Voltage +24V
Measurements without holder
Diameter×Length 102 × 150 mm
Holder radius of action ca. 470 mm
Weight ca. 3.2 kg

ROTAFLOW Emergency Unit

Speed display
Range 1500 to 5000 RPM
Resolution 500 RPM
Accuracy ±250 RP
Measurements without holder
Diameter×Length 102 × 270 mm
Length (folded handle) 185 mm
Other
Coupling magnetic
Gear 1:39
Holder radius of action ca. 210 mm
Weight ca. 2.2 kg

ROTAFLOW Centrifugal Pump

Pressure max. 750 mmHg
Flow 0 to 10 LPM
Connectors 3/8"
Material polycarbonate
Rotor diameter 50 mm
Surface 190 cm2
Priming volume 32 ml

How to order