Engineered with simplicity and sophistication, the ROTAFLOW Centrifugal Pump is one of the smallest centrifugal pumps available. It is a highly flexible modular component for systems integration.

The ROTAFLOW Centrifugal Pump is the only centrifugal pump without a metal shaft or seal. This pump features a peg-top, one-point sapphire bearing that lowers friction substantially. Its spiral housing ensures an optimized flow ratio with no stagnant zones. The Centrifugal Pump's minimal priming volume of 32 ml considerably reduces hemodilution.

The ROTAFLOW pump head is the perfect choice - reliable and efficient for standard procedures as well as high-performance situations.

The ROTAFLOW Centrifugal Pump features:

  • Minimal priming volume
  • Minimal surface area
  • No stagnant blood zones and less hot spots
  • Minimal hemolytic index
  • Minimal MTT (Mean Transit Time)
  • Hydraulic optimized flow channels
  • Spiral housing for best flow ratio
  • Low-friction one-point bearing (sapphire ball & PE calotte)

Technical Data

Safety
Certificate CE 0413
Standards IEC 60601-1, Type B, Class I
EMI EN 60601-1-2
Classification (MDD, Annex IX) IIb
Water protection IP64
Maximum flow deviation in case of error +20 %

Speed
Range 0 to 5000 RPM
Display resolution 1 RPM
Display accuracy ±20 RPM

Flow
Range 0 to 9.99 LPM
Display resolution 0.01 LPM
Display accuracy 0 to 1 LPM 0.07 LPM + offset drift
Display accuracy 1 to 9.99 LPM 7 % of actual value + offset drift
Offset drift max. 0.05 LPM

Environmental conditions
  Operation Storage Transportation
Temperature 17°C to 40°C -18°C to 45°C -18°C to 55°C
Relative humidity (non condensing) 30 % to 75 % 10 % to 96 % 10 % to 96 %
Air pressure 87 to 106 kPa 70 to 106 kPa 70 to 106 kPa

ROTAFLOW Console

Independent power supply, stand-alone
Mains (factory selectable)
100 V~ / 115 V~ / 230 V~ / 240 V~; 50 Hz / 60 Hz
Fuse
2 A for 100 V and 115 V 1 A for 230 V and 240 V
   
Power supply, Heart Lung Machine console
Voltage
+24 V
Current max.
8 A

Batteries
Type
NiCd
Voltage
24 V
Capacity
5 Ah
Charge/discharge cycles
min. 500
Shelf life
max. 6 months (under storage conditions)

SERial data output
Transfer rate
9600 baud
Protocol
8 N 1
other
Measurements, L×W×H    
179 × 385 × 243 mm
Weight ca.
14.4 kg

ROTAFLOW Drive Unit

Other
Coupling
magnetic
Speed
0 to 5000 RPM
Pressure max. 750 mmHg
Flow meter
ultrasonic
Bubble sensor
ultrasonic
Pump stop
air bubbles > 5 mm (0.065 cm3)
Motor
brushless DC
Voltage    
+24V

Measurements without holder
Diameter×Length
102 × 150 mm
Holder radius of action
ca. 470 mm
Weight
ca. 3.2 kg

ROTAFLOW Emergency Unit

Speed display
Coupling
magnetic
Gear
1:39
Holder radius of action
ca. 210 mm
Weight ca. 2.2 kg

Measurements without holder
Diameter×Length
102 × 270 mm
Length (folded handle)
185 mm

Other
Coupling
magnetic
Gear  
1:39
Holder radius of action
ca. 210 mm
Weight
ca. 2.2 kg

ROTAFLOW Centrifugal Pump

 
Pressure
max. 750 mmHg
Flow
0 to 10 LPM
Connectors  
3/8"
Material    
polycarbonate
Rotor diameter
50 mm
Surface 
190 cm2
Priming volume
32 ml

The ROTAFLOW Centrifugal Pump is a component of the extracorporeal circuit and can be used in conjunction with the ROTAFLOW Console.

Safety
Certificate CE 0413
Standards IEC 60601-1, Type B, Class I
EMI EN 60601-1-2
Classification (MDD, Annex IX) IIb
Water protection IP64
Maximum flow deviation in case of error +20 %
Speed
Range 0 to 5000 RPM
Display resolution 1 RPM
Display accuracy ±20 RPM
Flow
Range 0 to 9.99 LPM
Display resolution 0.01 LPM
Display accuracy 0 to 1 LPM 0.07 LPM + offset drift
Display accuracy 1 to 9.99 LPM 7 % of actual value + offset drift
Offset drift max. 0.05 LPM
Environmental conditions
  Operation Storage Transportation
Temperature 17°C to 40°C -18°C to 45°C -18°C to 55°C
Relative humidity (non condensing) 30 % to 75 % 10 % to 96 % 10 % to 96 %
Air pressure 87 to 106 kPa 70 to 106 kPa 70 to 106 kPa

ROTAFLOW Console

Independent power supply, stand-alone
Mains (factory selectable) 100 V~ / 115 V~ / 230 V~ / 240 V~; 50 Hz / 60 Hz
Fuse  2 A for 100 V and 115 V 1 A for 230 V and 240 V
   
Power supply, Heart Lung Machine console
Voltage +24 V
Current max. 8 A
Batteries
Type NiCd
Voltage 24 V
Capacity  5 Ah
Charge/discharge cycles min. 500
Shelf life  max. 6 months (under storage conditions)
SERial data output
Transfer rate 9600 baud
Protocol 8 N 1
other
Measurements, L×W×H      179 × 385 × 243 mm
Weight ca. 14.4 kg

ROTAFLOW Drive Unit

Other
Coupling magnetic
Speed  0 to 5000 RPM
Pressure max. 750 mmHg
Flow meter  ultrasonic
Bubble sensor ultrasonic
Pump stop air bubbles > 5 mm (0.065 cm3)
Motor  brushless DC
Voltage      +24V
Measurements without holder
Diameter×Length 102 × 150 mm
Holder radius of action ca. 470 mm
Weight ca. 3.2 kg

ROTAFLOW Emergency Unit

Speed display
Coupling magnetic
Gear 1:39
Holder radius of action ca. 210 mm
Weight ca. 2.2 kg
Measurements without holder
Diameter×Length 102 × 270 mm
Length (folded handle) 185 mm
Other
Coupling magnetic
Gear    1:39
Holder radius of action ca. 210 mm
Weight ca. 2.2 kg

ROTAFLOW Centrifugal Pump

 
Pressure max. 750 mmHg
Flow 0 to 10 LPM
Connectors    3/8"
Material      polycarbonate
Rotor diameter 50 mm
Surface   190 cm2
Priming volume 32 ml

How to order

To place an order or learn more about any of the Maquet products, related accessories, or therapeutic solutions, please contact your local sales representative. 

Related Information